Electrospray
helpful hints and answers to FAQs.
Some
general information that everyone should have…
1. QQQ and LC-TOF basic plumbing. The two LC systems - on the QQQ and the
ESI-TOF - are more different than they look.
The one on the triple-quad is usually configured for a maximum flow-rate
of 20 μL/minute, and
should only be used with columns 1mm ID or less. The LC on the TOF can be used routinely with
standard columns (2 - 4 mm ID, but I prefer 2 mm). This can also be used with capillary columns,
but a splitter and manual injector are required except for isocratic
analyses. The reason for this
requirement is that gradients won’t reach the column for a LONG time at the low
flow rates (a few μL/minute or even less) used
with narrow-bore columns. Another
consideration is that there are three flavors of sprayers, ‘micro’ for
capillary columns and flow-rates up to about 20 μL/minute,
‘standard’ for flow-rates up to about 200 μL/minute,
and ‘standard-with-spacer’ for higher flow-rates. All of these will at least work under all
conditions (although the ‘micro’ will develop high back-pressure at the higher
flows), but the spray assembly should match your flow rate for optimum
performance. (Don’t try changing the
sprayer, though, unless I’ve checked you out on the procedure.) The QQQ is usually set up with the micro
sprayer, but the configuration on the ESI-TOF is changed frequently, so if
you’re having sensitivity issues, check to see which sprayer is being used.
2. How little can I inject? The autosampler on
the triple-quad can reliably inject as little as 200 nL,
but the one on the TOF is not reproducible at those levels.
3. How much can I inject? In this same vein, injection volumes should
be based on the flow rate. If, for example,
you’re using a narrow-bore column at a flow-rate of 3 μL/minute, you should probably inject less than
1 μL; if you’re on the TOF, with a flow-rate of
200 μL/minute, you could probably go as high as
25 - 30 μL.
4. Fragile UV
flow-cells. The UV cell on the TOF can only handle pressures
below 60 bar.
So you should pay continual attention to your analyses because
the spray capillary can occasionally clog, and this can quickly develop enough
back-pressure to destroy the UV cell (replacements cost over $1000.00).
5. Basic buffer guidelines. Use the minimum buffer concentration that
will give you adequate chromatography, and use only volatile buffers. Even volatile buffers, e.g., ammonium
acetate, can suppress ionization and can thus cause problems if you have low analyte concentrations.
6. Be gentle. Try to never inject crude reaction mixtures
(I know it’s tempting), especially if there is a small amount of product in the
presence of huge amounts of starting material or - worse - salts. Protein digests should be zip-tipped before
injection. The triple-quad should be
able to routinely detect low-to- mid-femtomoles of
pure material (on-column) and the TOF somewhat less, so there is usually no
good reason to inject more than a picomole or so (and
some good reasons not to inject more, e.g., the ion source gets dirty faster or
the chromatography may deteriorate).
Start with a dilute sample and increase the concentration if necessary,
rather than vice-versa. Also, see
the next section.
7. How accurate is ‘accurate’? If you’re using the ESI-TOF for accurate mass,
be sure to check the calibration. I
check and calibrate frequently, and the instrument is fairly stable, so the
mass accuracy is usually within specifications, i.e., 3 ppm or better. But ‘usually’ does not mean ‘always’, so if
mass accuracy is important, you should take the time to check the
calibration. In addition, the instrument
often calibrates at far better than 3 ppm, so the extra effort will usually get
you better data. If you need the best
accuracy that the instrument can deliver, you should use the reference sprayer
(although there is often enough residual calibrant to
use as an internal reference). The
detector on this instrument will begin to saturate at ion-counts near or above
107, leading to sometimes-wildly inaccurate masses.
8. Why should I use a small column? Remember that electrospray is
concentration-dependent and flow-rate/droplet-size dependent. For a given amount of material, you will get
maximum sensitivity with the smallest column diameter, lowest flow-rate and the
smallest spray aperture. The downside is
that the smaller you go the more technical problems you encounter. Semi-micro or micro columns
- 1 mm to 0.35 mm ID with flow rates in the 10 - 3 μL/minute
range on the triple-quad with the micro spray offer a nice compromise, since
these are easy to work with and typically give excellent results.
9. Will I see anything at all? Always run something familiar before running
a real sample. This may seem too
cumbersome - especially if you’re in a hurry - but will definitely save you
time in the long run.
10. Those dangerous
first few minutes. Never leave the instrument during the first 10
minutes or so of an analysis that you’ve just set up. Stay around long enough to make sure that the
flow is stable, that the injection completes, and that no leaks develop. Check back frequently to make sure that
everything remains ok. If you’re running
a long work-list, make sure the first two or three samples run smoothly before
leaving the lab. Don’t push ‘Start’ and
walk away.
111. Is the spray working? Before starting an analysis, make sure that
the mass spectrometer signal is stable and robust. You should always see some background ions with
nice isotope peaks. If the background
signal is weak and noisy, with no stable peaks, you should deal with this before
running a sample. When in doubt, go to a
low mass range, e.g., 80 - 600; there are always solvent contaminants in this
range that show up well if the instrument is working properly. The most common cause of poor signal is low
flow (e.g., a leak) or inappropriate spray conditions, although a dirty
spray-shield can also be the villain.
13. Optimizing the spray. The four things most often adjusted to
optimize the spray itself are 1. nebulizing-gas
pressure; 2. drying gas flow; 3. drying
gas temperature; and 4. spray voltage. The actual values depend mainly on solvent
composition and solvent flow but also to some extent on the analyte. Drying gas temperature can be a problem with
low flows and concentrated samples; if it’s too high, the solvent may evaporate
too quickly from the spray capillary and precipitate sample, leading to a
plugged capillary (these are almost impossible to clean and cost over $100.00
to replace).
13. What should I use to dissolve my sample? The solvent that you use for the autosampler should closely match the initial solvent
composition for your gradient. If there
is a significant mismatch, you may get bizarre chromatography or - worse - the
sample may precipitate out and clog the injector needle or the switching valve.
14. Autosampler
inserts. You should generally use
glass inserts rather than plastic inserts in the autosamplers
in order to minimize plasticizer contaminants, and always use Agilent vials. Never put tape labels on the vials since this
can prevent them from going completely into the tray; this can halt the
injection sequence.
15. Liquid nitrogen. Always keep an eye on the supply of liquid
nitrogen. The indicators on the tanks
are notoriously unreliable, so you may have to shake the dewar to see how heavy it is. The nitrogen pressure for the ion triple-quad
should be 80 psi or slightly above; for the TOF it should be around 90
psi. The high-pressure setting on the
regulator is often confusing, since it usually reads just above the peg. Most people are familiar with gas cylinders,
which often have pressures above 2000 psi when fresh. The liquid nitrogen dewars in the lab, however, are set for about 240 psi
as head pressure, which barely moves the needle on a standard regulator, but
which is just right for the instruments.
When you replace a tank, notify Marcia so she can keep us supplied.
16. Remote data analysis. (Not strictly electrospray,
but useful.) The new Dell
computer in 56-747A has data analysis software for the QStar,
the ESI-TOF, and the triple-quad, along with ChemDraw
and Microsoft Office. The electrospray
mass spectrometers and two of the computers in 56-747A are all on a local
network; the Spectrum Mill server is on the web but - for security reasons - is
not connected to the local network.
17. Tune files on the QQQ. The QQQ is more compatible than the TOF with
customized tune files since the analytical requirements vary widely depending
on the acquisition mode.
18. Stabilizing flow on the triple-quad. The Agilent capillary pumping systems need a
fair amount of backpressure to stabilize.
So - if you try to run the pumps at low flows (1 - 5 μL/minute)
without a column, they’ll overshoot the setpoint,
oscillate wildly, and then shut down. If
you need to troubleshoot without a column, set the flow to 20 μL/minute and it’ll work fine.
19. How much background should I see? This depends on lots of things, e.g., solvent
composition, solvent quality, spray parameters, and scan range. A typical/reasonable background on the QQQ
base-peak chromatogram would be ion counts in the low-to-mid 105
range with a semi-capillary column (0.35 mm - 1 mm), but this could go below 1
x 104 with nanospray. The TOF will give roughly comparable counts
(low-to-mid 105 in the total ion chromatogram) with a fairly clean
solvent at 200 μL/minute. But remember the first sentence of this
paragraph!