| 
 Raman spectroscopy for measurement of blood analytesBackgroundMeasurement of the concentrations of blood analytes presently 
              requires withdrawal of one of more blood samples and a measurement 
              process which often involves sample handling, such as serum extraction, 
              addition of various reagents and a delay in the diagnosis process. 
              Withdrawal of blood exposes personnel to biohazards and causes inconvenience 
              and pain to the patient. A non invasive measurement would revolutionize 
              medical diagnosis by providing analytes concentrations quickly, 
              painlessly and without the use of reagents. A non-invasive measurement 
              would be particularly beneficial where the results are needed quickly 
              or where measurements must be taken frequently. An obvious example 
              of this is the measurement of glucose concentration. Millions of 
              people with diabetes must measure their glucose level multiple times 
              per day to maintain their glucose level within prescribed limits 
              so as to reduce the serious long term consequences of this disease. 
              Non-invasive measurement of glucose is a goal of many institutions. 
              Many technologies are being investigated to reach this goal. Among 
              them are absorption spectroscopy, both by diffuse reflectance and 
              transmission, light polarization and light scattering.
 Over the past several years, we have been investigating the use 
              of near-infrared (NIR) Raman Spectroscopy for measuring the concentrations 
              of blood analytes. The ultimate goal of this research is to develop 
              a transcutaneous method of measuring clinically important analytes 
              in the blood-tissue matrix by developing a basic understanding of 
              the scientific and engineering issues involved.  The strength of Raman spectroscopy lies in its sharp spectral features, 
              characteristic for each molecule. This strength is ideally suited 
              to blood analyte measurements, where there are many interfering 
              spectra, many of which are much stronger that that of blood analytes. 
              Raman’s sharp spectral features enable detection of blood 
              analyte spectra among these strong interfering spectra and in the 
              presence of a large fluorescence background. Figure 1 shows the 
              sharp spectral features of the glucose spectrum and how it is differentiated 
              from the spectra of the many components that exist in skin. 
               
                |  |   
                | Figure 1. The distinct spectral features, 
                  characteristic for a back-ground subtracted Raman spectrum, 
                  are here exemplified by a spectrum of glucose dissolved in water. 
                  The typical Raman spectrum of skin is shown for comparison with 
                  arrows indicating regions that clearly differentiate glucose 
                  from the other components in skin. To allow comparison, the 
                  glucose spectrum shown is 150 times the estimated size of the 
                  spectrum that would be received from a concentration of 5mM 
                  glucose in human skin. |  However, to measure glucose in tissue is far more complex than 
              indicated by this figure for various reasons: a) the spectra from 
              typical physiological concentrations of glucose in skin are in the 
              order of several hundred times lower than the total spectrum from 
              skin, as shown in an example in Figure 1, b) even with excitation 
              at 830nm, the fluorescence background dominates the signal, as can 
              be seen in figure 2, c) Raman peaks from all other biomolecules 
              that are present overlap and therefore interfere and d) the optical 
              properties of the tissue as well as the probe depth/volume influence 
              the signal.  
               
                |  |   
                | Figure 2. A signal collected from a transcutaneous 
                  measurement, is shown in blue. It consists of a dominating fluorescence 
                  background on top of which the Raman peaks from all present 
                  biomolecules is superimposed. The extracted Raman signal is 
                  shown in red. |  Previous ResearchPrevious work determined that the measurements of blood 
              analytes in a serum matrix were feasible. From that foundation, 
              the Raman system was improved and then utilized to demonstrate the 
              feasibility of the measurement of Glucose, Urea, Triglyceride, Total 
              Protein, Albumin, Hemoglobin and Hematocrit in whole blood [8]. 
              Based upon these successes, our focus moved to transcutaneous measurement 
              of analytes. To support this objective, the Raman system was further 
              improved to significantly increase its light collection and detection 
              efficiency. The system developed for this application is shown in 
              figure 3. The Raman light generated in the tissue is collected by 
              a paraboloidal mirror, designed for both wide-angle and large-area 
              light collection optimal for light being emitted from highly scattering 
              skin. A circular-to-linear shaped fiber bundle efficiently guides 
              the collected Raman light to the spectrograph and a large area CCD, 
              enabling recordings of Raman spectra with high sensitivity.
   
               
                |  |   
                | Figure 3. Diagram of the high sensitivity 
                  Raman spectroscopy system used for transcutaneous measurements |  Transcutaneous MeasurementsAs an initial evaluation of the ability of Raman spectroscopy 
              to measure glucose transcutaneously, a series of spectra were collected 
              on human volunteers in conjunction with a glucose tolerance test. 
              This involves the intake of a high-glucose containing fluid (SUN-DEX), 
              after which the glucose levels were elevated to more than twice 
              that found under fasting conditions. Raman spectra, each measured 
              for 3 minutes, and reference glucose concentrations from blood samples 
              were measured periodically during the 2-2 ½ hour duration 
              of the procedure for each volunteer. A Hemocue glucose analyzer 
              provided the reference measurement for the blood analysis. The manufacturer’s 
              specification for precision is SD < 6 mg/dl.
 The Raman spectra were extracted from the large fluorescence background 
              using a 5th order polynomial to fit the background. A calibration 
              algorithm was generated individually from the data from each volunteer 
              using the Partial Least Squares (PLS) regression method. Each calibration 
              algorithm was validated using hold-out-one cross validation.  A comparison of the predicted glucose concentrations to the corresponding 
              reference data from one of the volunteers is shown in figure 4. 
              The average error in the validated data (Standard Error of Validation, 
              SEV) is 9.8 mg/dl with an R^2 of 0.91.  
               
                |  |  |   
                | Figure 4. The left chart shows 
                  the predicted glucose tracking the reference values for one 
                  volunteer. The correlation of the same data is shown on the 
                  right, with an average error of 9.8mg/dl and an R^2 of 0.91. |  The procedure was applied individually to data from each of 16 
              volunteers and the validated prediction results combined into one 
              chart, shown in figure 5. For the data from all 16 volunteers, the 
              average prediction error is 13.2 mg/dl and the R^2 is 0.79.  
               
                |  |   
                | Figure 5. Cross validated results for 16 
                  volunteers calibrated individually. The average prediction error 
                  for this set is 13.2 mg/dl and the R2 is 0.79. |  A question that occurs with this kind of procedure is whether the 
              calibration is based upon glucose. Raman Spectroscopy offers a way 
              to obtain an indication of whether glucose is an important factor 
              in this calibration; by comparing the calibration regression vector 
              to the spectrum of glucose. Figure 6 compares the regression vector 
              for the calibration shown in figure 4 to the spectrum of glucose 
              in water, scaled to fit on the same chart. Due to the sharp features 
              of Raman Spectroscopy, there are numerous peaks in the glucose spectrum 
              that appear in the regression vector.  Unlike many methods of measuring glucose, where there are valid 
              questions of whether glucose is being measured, the existence of 
              numerous glucose peaks in the regression vector developed from Raman 
              measurements provides direct evidence that glucose is being measured. 
               
                |  |   
                | Figure 6. The regression (B) vector for the 
                  calibration shown in figure 4 and the spectrum of glucose, scaled 
                  to fit on the same chart. Numerous peaks in the glucose spectrum 
                  match peaks in the regression vector, indicating that glucose 
                  in an important part of the calibration. |  Current WorkThe data from this series of tests in volunteers has provided 
              a wealth of knowledge for us. We are continuing to analyze the data 
              to identify opportunities to improve results. From the analysis 
              and further testing of system characteristics, we have generated 
              a number of instrument improvements to be made and identified a 
              number of causes for error. Addressing those causes and making identified 
              instrument improvements are expected to decrease measurement error.
  We have further analyzed the optical and stability characteristics 
              of our system. Based upon that, we have improved light collection 
              efficiency by over 30% by changing to a higher NA fiber in the bundle 
              that couples light to the spectrometer. We are also taking steps 
              to make our system more stable and developing techniques to accurately 
              and precisely measure and correct for the drifts that remain.  We are also developing new data processing techniques to extract 
              more information from our measurements. Our goal is to utilize all 
              the techniques we are learning to obtain reduced error levels in 
              an expended human volunteer study.
 Recent Publications 
              "Measurement of blood analytes in turbid biological tissue 
                using near-infrared Raman spectroscopy", Tae-Woong Koo, Doctoral 
                thesis, Massachusetts Institute of Technology, 2001 "Prospects for In Vivo Raman Spectroscopy", Eugene 
                B. Hanlon, Ramasamy Manoharan, Tae-Woong Koo, Karen E. Shafer, 
                Jason T. Motz, Maryann Fitzmaurice, John R. Kramer, Irving Itzkan, 
                Ramachandra R. Dasari, and Michael S. Feld, Physics in Medicine 
                and Biology 45(2), R1-R59 (2000)"Reagentless Blood Analysis by Near-Infrared Raman Spectroscopy", 
                Tae-Woong Koo, Andrew J. Berger, Irving Itzkan, Gary Horowitz, 
                and Michael S. Feld, Diabetes Technology & Therapeutics 
                1(2), 153-157 (1999). "Multicomponent Blood Analysis by Near-Infrared Raman Spectroscopy", 
                Andrew J. Berger, Tae-Woong Koo, Irving Itzkan, Gary Horowitz, 
                and Michael S. Feld, Applied Optics 38(13), 
                2916-2926 (1999). "Measurement of Glucose in Human Blood Serum using Raman 
                Spectroscopy", Tae-Woong Koo, Andrew J. Berger, Irving Itzkan, 
                Gary Horowitz, and Michael S. Feld, IEEE-LEOS Newslette 12(2) 
                18 (1998). "An Enhanced Algorithm for Linear Multivariate Calibration", 
                Andrew J. Berger, Tae-Woong Koo, Irving Itzkan, and Michael S. 
                Feld, Analytical Chemistry 70(3), 623-627 
                (1998)."Measurements of analytes in whole blood by means of Raman 
                spectroscopy" Annika M. K. Enejder, Tae-Woong Koo, Jeankun 
                Oh, Gary L. Horowitz, Ramachandra R. Dasari, and Michael S. Feld, 
                SPIE's BIOS 2002, 19-25 January 2002, San Jose, California, 
                USA"Blood Analysis by Raman Spectroscopy" Annika M. K. 
                Enejder, Tae-Woong Koo, Jeankun Oh, Martin Hunter, Slobodan Sasic, 
                Gary L. Horowitz, Michael S. Feld. Optics Letters 27, 
                2004-2006 ( 2002). Support This research is funded by National Center for Research 
              Resources (National Institute of Health) and Bayer Diagnostics.
    
 |  |