|
In the closed
state (shown below) the scanner looks like a normal flip case. The user
would naturally press the latch button to open the “eyeglass case.”

When the
latch button is pressed, the two end caps pop out to reveal the scan button
on the front and a data/power port on the back.

The
reed switch is mounted at a fixed portion in the scanner while the magnet
is mounted on one of the sliding parts. When the magnet gets close to
the reed switch, it is triggered ON. This switch is used to turn on LEDs
for the instructional panel on the left.

The
latch button is mounted via the button flexure to the scanner case. The
flexure provides the restoring force for the button to bring it back to
its original state.
When
the latch button is pressed with the scanner in the closed state, it pushes
on the tip of the latch flexure that holds the two green sliding pieces
together. Springs are connected between the connection points (four springs
in total) add value of spring constant here on the case and the sliding
pieces.

When
the scanner is closed, the springs are in tension. As the latch flexure
tip is pushed out of the slot, the sliding pieces retract and the tip
of the latch flexure rides on the surface of the sliding part (see picture
below). The scanner can be closed by pushing the two end caps together.
As the scanner is closed, the latch flexure rides up onto the sliding
piece and continues to ride along it, until it falls into the slot.

(Roll
over image to see sliders open and close.)
|