Equipment
Lab and Coral Name | ICL / oxEtch-BOE |
Model | Semifab WPS-400 |
Specialist | Paul Tierney (Robert Bicchieri) |
Physical Location | 2F 3-South Photo |
Classification
Process Category | Wet |
Subcategory | Acids |
Material Keywords | SiO2, Acids |
Sample Size | 6" Wafers, 4" Wafers |
Alternative | TRL / acid-hood |
Keywords | multi wafer, manual load, both sides of sample, isotropic etch, manual operation |
Description
The oxEtch-BOE is an acid wet station that is dedicated for etching silicon dioxide. The buffered oxide wet etchant (7:1, NH4F:HF) has excellent selectivity to silicon and silicon nitride. The standing BOE bath is regularly aspirated and refilled by MTL staff.
Best for | Oxide stripping |
Limitations | Not suitable for samples that contain metal. |
Characteristics/FOM | 7:1 BOE |
Caution with | If bath is not at fill level, notify staff. Do not top off with water, as HF concentration will not be the same. |
Machine Charges | 25/tank |
Documents
SOP
BOE Etch | SOP for the ICL BOE etch bath |
Wet Processing | Rules and guidelines for wet processing |
Documents
Etch rates part 2 | Berkeley etch rate paper |
Process Matrix Details
Permitted
Germanium on surfaceSamples with germanium on the surface (typically grown films)
, Germanium buriedSamples with germanium buried below a different film
, Been in the STS DRIEThe DRIE etch leaves behind polymer residues on the sidewall ripples, which can be a contamination concern for some tools.
, Been in the SEMA sample viewed in the SEM must have used the appropriate chuck to avoid cross-contamination
, Been in the Concept1The Concep1 deposits dielectrics on GREEN wafers, however it also accepts metal and there can be cross-contamination for diffusion area
, Has PhotoresistSamples with photoresist cannot be exposed to high temperatures, which is typical in deposition tools. Outgassing can be a concern.
, Has PolyimidePolyimide is a very chemically resistant polymer, and can tolerate higher temperatures but cannot be exposed to typical PECVD deposition temperatures or diffusion furnaces. Outgassing can be a concern.
, Coming from KOHAfter a KOH etch, the samples must receive a special clean because the K ions are highly contaminating to CMOS frontend tools
, Coming from CMPAfter a CMP, the samples must receive a special clean, because the slurry residues otherwise introduce contamination and particles.
Not Allowed
Ever been in EMLSamples from EML are never permitted to return to ICL or TRL
, Been in the ALDSamples that have been in any of the ALD systems
, Pyrex SubstratesPyrex substrates can be a concern due to high sodium content, which contaminates CMOS frontend tools
, III-V SubstratesAny III-V substrates, e.g. GaAs, GaN, InP, and so on. Note though that many common III-V substrates will also carry the Au flag, but there are some GREEN III-V substrates.
, PiecesWafer pieces may not be handled by the equipment, and are harder to thoroughly clean - preventing them from running in certain tools.
, Gold or RED color codeRED color code substrates. These are gold-contaminated or have been processed in gold contaminated tools. Gold and other metals can contaminate silicon devices (GREEN color code) and have to be separated.
, Any exposure to CMOS metalIf the sample had ever seen a CMOS metal (or a tool that accepts CMOS metal), then some frontend tools could be contaminated by this.
, CMOS metal on surfaceCMOS compatible metals exposed on the surface. These are Al,Ni,Pt,Ti,TiN. Other metals such as Au are *NOT* part of this.
, CMOS metal buriedCMOS compatible metals covered entirely by a different material. These are Al,Ni,Pt,Ti,TiN. Other metals such as Au are *NOT* part of this.
, Has Cured SU8Not fully cured SU8 residues can heavily contaminated plasma chambers or destroy other user's samples, but fully cured SU8 is permitted in certain tools.
For more details or help, please consult PTC matrix, email ptc@mtl.mit.edu, or ask the research specialist (Paul Tierney)