Equipment
Lab and Coral Name | ICL / AME5000 |
Model | AMAT Precision 5000 |
Specialist | Eric Lim (Donal Jamieson) |
Physical Location | 2F 2-South DepAndEtch |
Classification
Process Category | Etch |
Subcategory | RIE |
Material Keywords | Silicon, SiO2, SiNx, Dielectrics |
Sample Size | 6" Wafers |
Alternative | ICL / LAM490B, ICL / LAM590-ICL |
Keywords | multi wafer, load lock, top side of sample, anisotropic etch, vacuum, plasma |
Description
The AME5000 is a reactive ion etcher (RIE) with RF generated, magnetically-coupled plasma. It has two chambers: Chamber A is designated for etching dielectrics, and Chamber B is designated for etching silicon and polysilicon. Chamber B can also serve as a backup for etching silicon nitride with SF6. This tool is GREEN and serves as the primary etcher for front-end processes and non-metal contaminated samples.
Best for | Etching silicon and dielectrics on GREEN samples, with straight smooth sidewall trenches, up to a few microns deep. |
Limitations | Not for deep etching of Si (use the DRIE tools instead) |
Characteristics/FOM | Chamber A gases:O2, CF4, CHF3; Chamber B gases: SF6, Cl2, HBr, O2 |
Caution with | Machine will start if cassette placed backwards (but will jam, don't do this!). Load wafers etch side up. |
Machine Charges | 7/wafer + 1/um |
Documents
SOP
AME5000 | SOP for the AME5000 etcher |
Documents
Etch rates part 2 | Berkeley etch rate paper |
Etch rates part 1 | Berkeley etch rate paper |
External Links
Dry Etching for VLSI | Fundamentals of dry etching, good introduction text |
Process Matrix Details
Permitted
Germanium on surfaceSamples with germanium on the surface (typically grown films)
, Germanium buriedSamples with germanium buried below a different film
, Any exposure to CMOS metalIf the sample had ever seen a CMOS metal (or a tool that accepts CMOS metal), then some frontend tools could be contaminated by this.
, CMOS metal buriedCMOS compatible metals covered entirely by a different material. These are Al,Ni,Pt,Ti,TiN. Other metals such as Au are *NOT* part of this.
, Been in the STS DRIEThe DRIE etch leaves behind polymer residues on the sidewall ripples, which can be a contamination concern for some tools.
, Been in the SEMA sample viewed in the SEM must have used the appropriate chuck to avoid cross-contamination
(With Appropriate Chuck), Been in the Concept1The Concep1 deposits dielectrics on GREEN wafers, however it also accepts metal and there can be cross-contamination for diffusion area
, Has PhotoresistSamples with photoresist cannot be exposed to high temperatures, which is typical in deposition tools. Outgassing can be a concern.
, Has PolyimidePolyimide is a very chemically resistant polymer, and can tolerate higher temperatures but cannot be exposed to typical PECVD deposition temperatures or diffusion furnaces. Outgassing can be a concern.
, Coming from KOHAfter a KOH etch, the samples must receive a special clean because the K ions are highly contaminating to CMOS frontend tools
, Coming from CMPAfter a CMP, the samples must receive a special clean, because the slurry residues otherwise introduce contamination and particles.
Not Allowed
Ever been in EMLSamples from EML are never permitted to return to ICL or TRL
, Been in the ALDSamples that have been in any of the ALD systems
, Pyrex SubstratesPyrex substrates can be a concern due to high sodium content, which contaminates CMOS frontend tools
, III-V SubstratesAny III-V substrates, e.g. GaAs, GaN, InP, and so on. Note though that many common III-V substrates will also carry the Au flag, but there are some GREEN III-V substrates.
, PiecesWafer pieces may not be handled by the equipment, and are harder to thoroughly clean - preventing them from running in certain tools.
, Gold or RED color codeRED color code substrates. These are gold-contaminated or have been processed in gold contaminated tools. Gold and other metals can contaminate silicon devices (GREEN color code) and have to be separated.
, CMOS metal on surfaceCMOS compatible metals exposed on the surface. These are Al,Ni,Pt,Ti,TiN. Other metals such as Au are *NOT* part of this.
, Has Cured SU8Not fully cured SU8 residues can heavily contaminated plasma chambers or destroy other user's samples, but fully cured SU8 is permitted in certain tools.
For more details or help, please consult PTC matrix, email ptc@mtl.mit.edu, or ask the research specialist (Eric Lim)